The advanced amplifier circuitry provides a frequency response from dc to
20 kHz with a corresponding noise level of 0.01 mV rms/ .
The Accumeasure 5000 can be used with any of MTI's standard or custom
designed miniature sensor probes for unmatched linearity, stability and
resolution. The Accumeasure System 5000 is available in either single or
dual channel designs. The dual channel model includes a summing amplifier
for applications that require sum or difference measurements.
APPLICATIONS
Spindle Runout
Displacement
Flatness |
Engine Dynamics
Positioning
Profiling |
Vibration
Structural Dynamics
Diameter Measurements |
FEATURES
|
High Frequency Response, up to 20 kHz for real time measurements. |
|
Non-contact Measurement with small probe sizes, including rectangular
and annular configurations. |
|
Offset and Gain Controls on front panel for ease of use and operation. |
|
Exceptional Stability for long term monitoring or experiments. |
SPECIFICATIONS:
| Linearity |
±0.1% over 10 to 100% of full-scale range using standard 8-ft probe cable |
| Output Signal |
0 to 10 V dc from rear-panel BNC connector
at 50-ohm impedance |
| Offset Control |
Allows standard output signal to be offset for -10 V to 0 V dc output |
| Probe Voltage |
Proportional to probe-to-target gap;
5V rms at full scale; 8V rms maximum |
Temperature Stability
|
(after 30-min warm up) 60°F - 95°F (16°C to 35°C), drift less than ±0.1% of range; 40°F
- 100°F (4°C to 4O°C), drift less than ±0.25% of range |
| Noise Levels |
1.7 mV rms, DC to 20 kHz;
0.5 mV rms, DC to 2 kHz;
0.011 mV rms/√Hz
0.1 µin. rms, DC to 2 kHz (50.8 µm probe)
0.003 µin. rms/√Hz (50.8µm probe) |
| Probe GAIN |
Standard; increased gain for greater probe range available upon request |
| Power |
120 Vac/240 Vac, 10 W 50/60 Hz |
| Operating Temperature |
40°F to 100°F (4°C to 40°C) |
Dimensions:
Width
|
7.5 in. (19.0 cm) |
|
Height |
2.5 in. (6.4 cm) |
| Depth |
11.25 in. (28.6 cm) |
| Weight |