THE MTI-2000 is a fiber optic noncontact vibration and displacement sensor with
exceptional resolution, high frequency response and a
choice of 11 plug-in probe modules. With frequency
response up to 200 kHz, it is ideal for precision noncontact sensing measurements. Fiber optic probes are
immune to electromagnetic interference and will operate on
almost any type of surface or material. Dual channel
capability makes simultaneous measurements possible.
General
Specifications
Power Requirements : 117
Vac @ 0.35 A or
220 Vac @
0.2 A, 50 to 400 Hz
Max Power Dissipation : 20 W
Line Voltage Stability
:105 to 130 Vac
Fuse : 0.5
A, fast-blow (requires IEC
3-pin grounded AC plug)
Meter
Readout : 4.5-digit green vacuum
fluorescent,
updates 4 times
per second
Bargraph
: 20-element green LE (0.5V
step size),
10 Hz response
Dimensions :
4.5 in. H x
13.3
in. W x 10.0 in D
Displacement
Measurements
Output Signal : 0 to ±10 V
dc
Output impedance :100 ohm
Switchable Offset Range : 0 to -15 V
Stability at 12 hrs.±2° F (-1°C) :
Drift < 1.0% of full
scale
Stability at 60°- 95° F (16-35°C):
Drift < 2.0% of full scale
Vibration
Measurements
Output Signal : 0 to ±10 Vdc full-scale
range,
100 ohm output impedance
Accuracy :
Within 1.0% for peak-to-peak
readings from 1 Hz-150 kHz
System
Noise : Dependent upon probe type
and
surface reflectivity
Environmental
Requirements
Operating Temperature : 50ºF- 110ºF (10ºC- 43ºC)
Storage Temperature : 0ºF-150ºF (-18º C to 65º C)
Weight
: 18 lb (6.1 kg)
Probe
Specifications
Temperature Range : -100°F- 300°F (-70°C-150°C)
Operating Pressure
:
Vacuum of 29 in. Hg- 500 psig
Tip Length : 3 in. (76.2 mm]
Cable Length : 54 in. (1372 mm] standard
Options
220 Vac input power configuration.
Display in metric engineering units
RS-232 digital output
Bent probe tips
High-pressure/ High-vacuum probes
Probe operating temp. from
-310°F-1382°F (-190°C-750°C)
Accessories
Optical Probe Extenders: Precision lens systems that
permit operation of Standard
probes at a greater
standoff
distance
with no loss of
probe
sensitivity or frequency
response
Fixture
Stands : Provides a stationary mounting
for probes of any size.
Calibration
Fixture : Secures the probe
and
calibrates
the instrument to a non-rotating
target, utilizing a precision
micrometer.
Used for initial
calibration, to a
fluid medium, or
in- place calibration
checks.
|