HOME

Search

Contact info.

  Data Acquisition
  Capacitance sensors
  
 MP402 pressure transmitter- 4-20 mA o/p:

GENERAL

Model MP402 is a miniature, low cost, general use pressure sensor that automatically compensates for line resistance and power supply changes. It is made of corrosive resistant titanium (6AL4V) and utilizes high reliability bulk type semi-conductor strain gages bonded directly to the inside surface of the machined diaphragm.


 
Model MP402
  4-20 mA O/P


Absolute or sealed reference sensors are available as an option. The hermetic header is located just behind the strain gage diaphragm section of the sensor, sealing the sensitive gages to an inert environment which dramatically increases stability and reliability.

Special temperature or pressure ranges and high-vibration resistant options including long term stability versions are available.

APPLICATIONS

  • TEST & MEASUREMENT EQUIPMENT
  • PRODUCTION EQUIPMENT  & CONTROLS
  • ENERGY CONTROLS
  • LABORATORY INSTRUMENTS   & EQUIPMENT
  • LIQUID LEVEL
  • CALIBRATION EQUIPMENT

MECHANICAL

Standard Pressure Ranges 25, 50, 100, 200, 300, 500, 1K, 2K, 3K, 5K, 7K, 10K PSI
Non Standard Pressure Ranges 5, 10, 15, 20, 30, 150 PSI
Other ranges possible Consult Factory
Over Pressure (No change out of spec.) 2.0x Range
Usable Pressure (Small change in balance) 3.0x Range
Burst Pressure 6.0x Range

PERFORMANCE OPTIONS (ALL VALUES MAXIMUM)


A B C D Units
Balance (Zero) 4.00
±0.50
4.00
±0.30
4.00
±0.20
4.00
±0.20
mA
Full Scale Sensitivity 16.00
±0.50
16.00
±0.30
16.00
±0.20
16.00
±0.20
mA
Static Error Band ±0.50 ±0.25 ±0.25 ±0.25 %FS
Thermal Balance (Zero) Shift ±0.02 ±0.01 ±0.01 ±0.005 %FS/°F
Thermal Sensitivity Shift ±0.02 ±0.01 ±0.01 ±0.005 %FS/°F
Long Term Stability ±2.00 ±2.00 ±2.00 ±2.00 %FS/year
*BFSL=Best Fit Straight Line


ENVIRONMENTAL

Storage and Operating Temperature 0° -180°F
Compensated Temperature Range 30° - 130°F
Acceleration (Consult factory for high vibration options) 100gÕs any axis
Media Compatibility (wetted surface) Liquids & Gases compatible with Titanium 6AL4V
Other Temperature Compensated Ranges Consult Factory

ELECTRICAL
 
Input Voltage (Standard) 24 VDC
Input Load Resistance (Standard) 500 Ohms
Maximum Input Voltage 30 V for short periods
Insulation Resistance Minimum 50 MOhms @ 50 VDC

 

Toll Free: (888) 738-0039
E-mail: a-kast@on.aibn.com

Copyright © 2000 A-KAST Measurements & Control Limited